EBV tests performance on DiaSorin’s Newly Launched LIAISON® XL

Mira Barak ,Rasha Daniel, Maya Grinberg, Clara Henig, Orit Rozenberg
Central Laboratory of Haifa and Western Galilee, Clalit Health Services. Ha'taasia 47, Nesher 20300, Israel

 

Objective: The aim of the present study was to evaluate the EBV (EBNA, VCA-IGG and IGM) kits panel on the LIAISON® XL, DiaSorin’s next generation analyzer, against the performance of the well established LIAISON® platform.

Methods: 120 routine samples were collected and used in the evaluation for the EBNA performance while 112 and 98 were assayed also for VCA-IGG and IGM. A proportion of 40%, 20% and 60% the samples chosen were below or around EBNA, VCA-IGG and IGM cut-offs, respectively. Samples were run on the LIAISON® and LIAISON® XL analyzers, on the same day. We checked the concordance between the two analyzers, and discrepant samples were rechecked in another lab or confirmed by blotting.

Results: The overall correlation between the two analyzers results was 96.9%. The concordance in the EBNA assay was 94.6% Among the 7 discrepant results it seems that the LIAISON XL had 2 FP and 4 FN and one result was inconclusive. The concordance in the VCA-IGG assay was 98.2%.  Among the 3 discrepant results it seems that the LIAISON had 1 FP, 1 FN and one result was inconclusive. The concordance in the IGM assay was 96.9%. Among the 3 discrepant results it seems that the LIAISON XL had 3 FP. In view of the algorithm adopted at our institution (initially performing EBNA and only if negative continuing to perform VCA-IGG and IGM), most of the discrepant results (borderline levels) will not affect the diagnosis. However, adjusting the lower cut-off of the EBNA assay to 6 U/ml instead of 5 U/ml renders similar clinical outcome between the analyzers. Using fine tuning of cut-offs, the overall correlation between the two analyzers was 98.8%.

Conclusion: The LIAISON® XL offers higher throughput, long walk-away time and improved instrument reliability. The performance of the EBV panel was comparable between the two platforms.

 

לתכנית הכנס     חזרה לריכוז הפוסטרים 2012


דף הבית | מי אנחנו | רישום לאילמ"ר ולאתר | תקנון | חברי הועד | מערכת האתר | צור קשר | מקצוע המעבדנות הרפואית | מישיבות ועד אילמ"ר | המועצה למעבדות במשה"ב |
חיפוש פרטי חבר | עדכון פרטים | ועדות מקצועיות | מפגשים וועדות המטולוגיה וקרישה | פורום אנדוקריני | פורום האיכות | כנסים שנתיים | ימי עיון | קפה מדע | כנסים של אגודות אחרות | פעילות בינלאומית | אבטחת איכות סקירת ספרות | מכשור ושיטות | דיווחי המעבדות | פרסים | לזכר חברנו | קישורים | דרושים | מידע מקצועי מחברות תומכות    