CLINICAL EVALUATION OF A NEW SPECIFIC 1-84 PTH ASSAY ON AUTOMATED PLATFORM

Mira Manor, Yitzhak Mayer
Endocrinology lab, Atidim Central lab, Clalit Health Care
12 Kiryat Atidim st. P.O.B 58150 Tel-Aviv 61581, Israel

 

Introduction: 2nd generation “Intact” PTH assays currently available, measure 1-84 PTH and additional fragments. Recently a third generation kit developed which detects only active 1-84 PTH molecule, based on an immuno-radiometric sandwich assay.

Methods: Evaluation took place in an HMO central lab. 240 plasma samples collected. 161 randomly out patients and 78 hospitalized nephrologic patients. Patients’ samples, controls and NEQAS samples tested by CLIA Method: DiaSorin LIAISON® 1-84 PTH assay, DiaSorin LIAISON® N-TACT® PTH II, SIEMENS Centaur iPTH and DiaSorin IRMA N-tact® PTH SP ("Gold Standard"). Results were compared with laboratory parameters and patients clinical status.

Results: Correlation between LIAISON® 1-84 PTH and LIAISON® N-TACT PTH II showed R= 0.9164, slope= 0.9076 . Correlation between LIAISON® 1-84 PTH and Centaur iPTH showed R= 0.98, slope= 2. 7622. Correlation between LIAISON® 1-84 PTH and IRMA N-tact® PTH SP showed R= 0.97, slope= 1.0893 .The following LIAISON® 1-84 PTH parameters obtained: Samples and controls within run precision showed  2.58%-5.34% CV. Samples and controls between run precision showed 4.03%-4.65% CV. Functional sensitivity matched manufacturer claim at 4 pg/mL. Recovery between 77%-101%. Clinical agreement between LIAISON® 1-84 PTH and IRMA N-tact® PTH SP on nephrologic patients was 97.5%. 3 discrepant patients were clinical matched with LIAISON® 1-84 PTH results. From the healthy out patients tested samples 54% were found high with SIEMENS Centaur iPTH and 16% were high with LIAISON® 1-84 PTH. The same behavior observed with NEQAS data. PTH levels were influenced similarly by low CA levels.

Conclusions: High diagnostic correlation was observed between IRMA N-tact® PTH SP and LIAISON® 1-84 PTH together with high clinical correlation of patients’ status. It is concluded that the LIAISON® 1-84 PTH can be used as a reliable and accurate kit for setting up a clinical and high throughput laboratory.

 

לתכנית הכנס     חזרה לריכוז הפוסטרים 2012
דף הבית | מי אנחנו | רישום לאילמ"ר ולאתר | תקנון | חברי הועד | מערכת האתר | צור קשר | מקצוע המעבדנות הרפואית | מישיבות ועד אילמ"ר | המועצה למעבדות במשה"ב |
חיפוש פרטי חבר | עדכון פרטים | ועדות מקצועיות | מפגשים וועדות המטולוגיה וקרישה | פורום אנדוקריני | פורום האיכות | כנסים שנתיים | ימי עיון | קפה מדע | כנסים של אגודות אחרות | פעילות בינלאומית | אבטחת איכות סקירת ספרות | מכשור ושיטות | דיווחי המעבדות | פרסים | לזכר חברנו | קישורים | דרושים | מידע מקצועי מחברות תומכות    